Technical Service

What is the operating procedure for Xiangtan tongyong centrifuges?


summary : This machine achieves clarification by separating solid impurities from liquids. Used to separate various difficult-to-separate suspensions. Especially suitable for solid-liquid separation with low concentration, high viscosity, fine solid particles, and small solid-liquid specific gravity. The solid phase should be removed after shutdown. So, what is the operating method of Xiangtan tongyong Centrifuge?

Xiangtan tongyong Centrifuge What is the operating method?

The machine achieves clarification by separating solid impurities from liquids. Used to separate various difficult-to-separate suspensions. Especially suitable for solid-liquid separation with low concentration, high viscosity, fine solid particles, and small solid-liquid specific gravity. The solid phase should be removed after shutdown. So Xiangtan tongyong Centrifuge What is the operating method?

 Chongqing Jiangbei centrifuge

Xiangtan tongyong Centrifuge Operating method:

1. When using various centrifuges, the centrifuge tubes and their contents must be accurately balanced on a balance beforehand. The weight difference during balancing must not exceed the range specified in the instruction manual of each centrifuge. Each centrifuge's different rotors have their own allowable differences, and an odd number of test tubes must not be loaded in the rotor. When the rotor is only partially loaded, the tubes must be placed symmetrically in the rotor so that the load is evenly distributed around the rotor.

2. If centrifugation is to be performed at a temperature below room temperature. Before use, the rotor should be pre-cooled in a refrigerator or centrifuge rotor chamber.

3. Do not leave during centrifugation. Observe whether the instruments on the centrifuge are working normally at all times. If there is an abnormal sound, stop the machine immediately for inspection and promptly troubleshoot.

4. Each rotor has its own maximum allowable speed and cumulative time limit. When using the rotor, please refer to the instruction manual and do not use it too quickly. Each rotor must have a usage document to record the cumulative usage time. If the maximum usage time limit of the rotor is exceeded, it must be used at a reduced speed as specified.

5. When loading the solution, it should be done according to the specific operating instructions of various centrifuges. According to the nature and volume of the liquid to be centrifuged, an appropriate centrifuge tube should be selected. Some centrifuge tubes are uncapped, and the liquid cannot be filled too much to prevent the rotating head from being thrown out during centrifugation, causing imbalance, rust, or corrosion.

Xiangtan tongyong Centrifuge What is the daily maintenance?

1. Rotating drum

Before operating the centrifuge, please cut off the power and release the centrifuge brake. You can try manually rotating the drum to see if there is any bite.

2. Check if other parts are loose and abnormal.

3. Turn on the power and start the car clockwise (usually it takes about 40-60 seconds from standstill to normal operation). Usually, each equipment must be idled for about 3 hours after leaving the factory, and it can only work without any abnormal situations.

4. Materials should be placed as evenly as possible.

5. Must be operated by a designated person, and the capacity must not exceed the rated capacity.

6. It is strictly forbidden to operate the machine at an overspeed, so as not to affect the service life of the machine.

7. After the machine starts, if there is any abnormal situation, it must be stopped for inspection, and disassembly, cleaning, and repair should be carried out if necessary. The centrifuge rotates very fast, so you must not touch its drum with your body to prevent accidents.

8. The mesh number of the filter cloth should be determined according to the size of the solid particles of the material to be separated, otherwise it will affect the separation effect. In addition, the filter cloth must also be installed.

9. The sealing ring is embedded in the sealing groove of the drum to prevent material from entering it.

10. To ensure the normal operation of the centrifuge, the rotating parts should be lubricated and maintained every six months. At the same time, check whether the bearing operation lubrication is worn; whether the parts in the braking device are worn, and replace them carefully; whether the bearing is leaking oil.

11. After using the machine, it should be cleaned and kept clean.