How to remove a centrifuge rotor that is stuck? A fully automatic centrifuge manufacturer teaches you.
summary :
What to do if the automatic centrifuge rotor cannot be removed? An automatic centrifuge manufacturer teaches you.
How to remove a centrifuge rotor that is stuck in a fully automatic centrifuge, as explained by a manufacturer?
Improper use of a fully automatic centrifuge can cause the rotor to become stuck, delaying the process. Generally, the rotor cannot be removed from the centrifuge chamber, mainly due to adhesion between the rotor and the centrifuge motor shaft. Below, Zhangjiagang Runxing Machinery, a manufacturer of fully automatic centrifuges, will guide you through several solutions. Details are as follows:
1. Simple Method
Unscrew the original locking screw of the fully automatic centrifuge and insert a screw with the same thread specification into the motor shaft thread hole. Note: Do not fully insert the screw. Two people cooperate; one person holds the rotor of the fully automatic centrifuge and gently lifts it upwards. Be careful not to use excessive force to avoid deformation of the motor support frame. The other person uses a hammer and a thin rod to tap the screw on the upper part of the motor shaft. After repeating this several times, the rotor can usually be separated from the shaft.
2. Specialized Tool Method
If the simple method fails to remove the rotor, it indicates that the adhesion is severe. Rust remover can be applied to the joint between the shaft and the rotor for rust removal and lubrication. After waiting for about a day, use a special puller to remove the rotor. Similarly, select a puller of the appropriate size according to the rotor size. Then, hook the puller handle onto the bottom of the rotor, and place the puller screw head against the screw in the motor shaft thread hole. After aligning the puller, rotate the screw clockwise with a wrench. According to the mechanical principle of the screw, the puller handle will generate a large pulling force, thereby separating the rotor of the fully automatic centrifuge from the shaft.
3. Important Points
1) Regardless of the method used, a replacement screw must be inserted into the motor shaft thread hole to protect the motor shaft thread and the original locking screw. Otherwise, damage to the original thread of the fully automatic centrifuge may result in motor failure.
2) Use appropriate force and avoid excessive force or hammering. If the resistance is too great, extend the rust removal and lubrication time.
3) After removing the rotor, use fine sandpaper to polish the surface of the motor shaft and the inner surface of the rotor to remove rust, and apply lubricating grease to prevent re-adhesion.
4. Preventive Measures
1) Strengthen daily maintenance. Regularly clean and lubricate the contact surface between the rotor and the shaft.
2) Especially for high-speed refrigerated centrifuges, do not close the door immediately after use. Allow the moisture, condensation, and corrosive gases in the centrifuge chamber to evaporate completely and return to normal temperature before closing the door.
3) Remove the rotor promptly after each centrifugation. If a rotor is not changed or removed for many days, it is very easy to cause adhesion. In severe cases, it may cause the entire machine to fail.
4) Do not overtighten the screws during tightening, otherwise, it may cause the screw to slip or detach, and in severe cases, it may cause motor failure. Because the motor usually rotates counterclockwise, when accelerating, due to inertia, the screw itself will generate a clockwise tightening force, which can tighten the rotor. Therefore, when tightening the rotor, it is only necessary to feel a slight effort.