summary :
Centrifuge Rotor Imbalance Solution
Solutions for Centrifuge Rotor Imbalance
It should be noted that other faults that cause large amplitude of radial vibration fundamental frequency of the centrifuge include: shaft misalignment, shaft bending, mechanical looseness and mechanical resonance. These should be distinguished from imbalance and corrected before checking for imbalance in the centrifuge.
If the 1X, 2X, 3X components are large, and the vertical vibration of the centrifuge is significantly greater than the horizontal vibration, it may be due to loose foundation.
If the axial vibration is large, and the radial and axial 1X, 2X, 3X components are large, it may be due to shaft misalignment.
Slightly changing the centrifuge speed, if the fundamental frequency amplitude changes significantly, it can be diagnosed as mechanical resonance.
For centrifuge motors, if the fundamental frequency amplitude is large and the vibration time-domain waveform shows slow modulation, it may be a motor fault, such as rotor breakage or cracks.
Shaft bending has similar spectral characteristics to imbalance. The method of distinction is: check the radial runout of each part of the rotor at low speed to determine if there is initial bending; change the unit load at a certain speed, if the vibration changes with the load and time, it may be local friction, uneven heating or cooling caused by thermal bending.
Contact Us
Company Fax: 0731-57884719
Company Address: Tanjiashan Town, Xiangtan City, Hunan
Chinese Website: Xiangtan Centrifuge. Website
English Website: http://www.tylxj.com