Technical Service
How to troubleshoot the instability of centrifuge speed?
summary :
How to troubleshoot the instability of centrifuge speed?
How to troubleshoot unstable centrifuge speed?
Fault phenomenon: The centrifuge is one of the indispensable important equipment in hospitals and research institutions. Its quality directly affects the centrifugation quality of the centrifuged substance. When I was repairing a centrifuge with unstable speed, I found that it was unstable at both high and low speeds.
Fault repair: This centrifuge is a DC speed-regulating motor. Its speed is controlled by changing the conduction angle of the thyristor through an adjustable resistor, and then the thyristor controls the rectified voltage reaching both ends of the motor to achieve the purpose of speed control. From the principle analysis, the following four situations are possible for unstable motor speed.
(1) Adjustable resistor: Because the adjustable resistor is installed on the control panel and has a very high usage frequency, it is common for poor contact to occur after a long time.
(2) There are also many problems with thyristors, usually breakdown short circuits, open circuits, inability to trigger, or internal poor contact.
(3) One of the four rectifier diodes has a problem.
(4) There is carbon powder between the motor rotor and the commutator, or the carbon brush itself is faulty. Following the principle of repairing easy problems first, I first measured the adjustable resistor. The resistance changed regularly with the adjustment, proving that the potentiometer was good. Further measurement of the thyristor also revealed no problems. When measuring the four rectifier diodes, one was found to be intermittently faulty. Since this diode is a large-current gold-sealed tube, one end is screwed to the circuit board, and the other end is welded to the circuit board. When removing it from the circuit board, loose positive and negative poles were found. While measuring, I shook it and confirmed that the diode was intermittently faulty. Because the model of the diode could not be clearly seen, I installed a tube of the same specification. After powering on, the speed was very fast, indicating that the original tube was broken down. After removing and measuring, it was found to be undamaged.
(5) The measurement of the other three diodes was also normal. However, during the comparative measurement, it was found that the positive and negative polarities of the two types of diodes were completely opposite (that is, the end with the screw has positive polarity, and some have negative polarity). Therefore, when installing, be sure to use components of the same model and specification. When components of the same specification and model were replaced, the power was turned on for testing, and everything returned to normal.