  Operation, Maintenance, and Troubleshooting of Centrifuges
 
  I. Operation and Maintenance of Centrifuges
 
  Centrifuges operate at high speeds; therefore, safety precautions must be observed during operation. The following points must be adhered to:
 
  1
 
  When using a centrifuge, place it on a stable and sturdy surface (its base is equipped with rubber suction feet, which, with the help of atmospheric pressure and the weight of the instrument itself, adhere firmly to the surface); large-capacity low-speed centrifuges and high-speed refrigerated centrifuges should be placed on a solid, level surface. High-speed centrifuges, weighing hundreds of kilograms, should be placed on a very solid surface with dust and moisture-proof equipment to ensure that the centrifuge chamber reaches a vacuum and operates normally. Before operating the centrifuge, the load must be balanced, with the weight difference as small as possible, otherwise, it will cause violent vibrations and damage the centrifuge rotor and shaft.
 
  2
 
  The centrifuge separation bracket must be in place before the centrifuge tubes are installed and operated. Empty operation is strictly prohibited. When starting the centrifuge, the speed switch or G-value knob should be set to a low position, and then gradually adjusted to the required value. After each operation, the knob must be adjusted to a low position before the next operation. Direct starting is prohibited at high-speed or high G-value settings.
 
  3
 
  Select the appropriate rotor and control the rotor speed. The high-speed rotation of the rotor should be reduced under the following circumstances: when the rotor's operating time and number of operations reach its specified lifespan; when the rotor is partially surface damaged or shows slight corrosion inside the tube holes; when using stainless steel centrifuge tubes, sleeves, thick-walled tubes, or centrifuge bottles; when the average density of the sample exceeds 1.2 g/cm3; when the rotor is subjected to high temperatures.
 
  4
 
  Before each use, carefully inspect the rotor for any foreign objects or dirt inside the holes to maintain balance. The rotor must be kept dry and clean and must not be bumped or scratched. If accidentally dropped, the rotor should be X-rayed to confirm that there is no internal damage before further use. After each use, the rotor should be cleaned by soaking in warm water (50℃~60℃) and neutral detergent, rinsed with distilled water or deionized water, wiped dry with a soft cloth, dried with a hair dryer, waxed, and stored dry.
 
  5
 
  The tensile strength and elongation of various centrifuge tubes are different, and should be selected as needed. Centrifuge tubes can be made of many different materials, such as polyethylene, cellulose, polycarbonate, polypropylene, etc. When using, ensure that the sample, solvent, and gradient material do not corrode the centrifuge tube and cap assembly. Expired, aged, cracked, or corroded centrifuge tubes should be avoided or used at reduced speeds. Control the number of times plastic centrifuge tubes are used and pay attention to matching specifications. Do not open the centrifuge chamber lid during centrifugation, and do not touch or allow foreign objects to collide with the rotating rotor and centrifuge tubes.
 
  6
 
  For low-temperature centrifugation samples, the empty rotor should be pre-cooled at 2000 rpm for a certain period of time. During pre-cooling, the temperature should be controlled at around 0℃. The rotor can also be pre-cooled in a refrigerator for several hours. Centrifuge cups can be directly stored in a refrigerator for pre-cooling.
 
  7
 
  If any abnormalities are found during centrifugation, such as unusual noise and vibration, stop the machine immediately and check. Do not continue operation until the cause is found.
 
  8
 
  During the daily use of the centrifuge, the level of the main unit should be calibrated every three months, the vacuum pump oil should be processed every 500 million revolutions, and the drive motor carbon brushes should be checked every 500 hours of operation. Remove the carbon powder generated by friction using a vacuum cleaner, and replace the carbon brushes or polish the commutator with fine sandpaper if necessary. The drive unit bearings should be cleaned and high-speed lubricating grease added approximately every 1500 hours of use. The joint between the shaft and the rotor should be regularly lubricated to prevent rust. When not in use for extended periods, apply rust-preventative oil and wrap with oil-impregnated paper. When not in use, the machine should be run at low speed for 1-2 times per month, each time for 0.5 hours, to ensure the normal operation of all parts.